Calibration Screening
Binary angular positioning routines establish a fundamental verification method for linear inertial transducers by using Earth’s gravity as a fixed input standard. The two position tumble test involves rotating an accelerometer through exactly one hundred and eighty degrees around a horizontal axis to place its sensitive axis in opposing vertical alignments. This maneuver subjects the internal sensing element to positive one g and negative one g reference accelerations.
The methodology ceases to be effective for sensors whose input range spans hundreds of gravities, where Earth’s field provides insufficient signal amplitude to evaluate high-range linearities.
Reversal Mechanism
Mechanical manipulation requires rotating the instrument along a rigid fixture indexed against physical stops or optical dividing heads. In the initial orientation, the sensitive axis aligns parallel to the gravity vector, driving maximum positive signal output from internal sensing structures. The technician or automated arm rotates the fixture one hundred and eighty degrees, inverting the transducer so that gravity acts in the reverse direction across the proof mass.
Both stationary outputs are recorded once dynamic ringing subsides, capturing paired electrical potentials corresponding to a precise two-g total input differential.
Parameter Isolation
Mathematical treatment of the two collected data values isolates basic sensor operational parameters. Subtracting the negative-orientation voltage from the positive-orientation voltage and dividing the difference by two gravities yields the precise scale factor of the transducer. Adding the two raw output voltages directly cancels out the true acceleration response, leaving exactly twice the zero-g bias offset voltage.
This simple extraction isolates offset error from sensitivity drift without requiring complicated multiaxial calibration fixtures or active dynamic centrifuges.
Inspection Limits
Acceptance testing in sensor manufacturing facilities uses this screening procedure to filter out assembly defects. Sourcing specifications mandate strict limits on permissible bias voltage and scale factor tolerance, rejecting parts that fail to meet datasheet baselines under the two-g inversion swing. Test fixtures require precise mechanical alignment, since a tilt error of one degree from true vertical introduces a cosine error that reduces effective gravitational acceleration, corrupting scale factor calculations.
Verified sensors receive factory calibration certificates detailing tested bias and sensitivity before packaging.