Temperature division
Defined mathematical interpolation equations apply to specific temperature intervals within the international temperature scale. The use of an its-90 sub-range allows calibration laboratories to qualify standard platinum resistance thermometers over restricted spans without performing the full set of fixed-point calibrations. These intervals are designed to meet the specific requirements of different industrial and scientific applications.
Mathematical formulation
Interpolation behaves according to specified reference functions and deviation equations defined for each interval. The thermometer is calibrated at a selection of fixed points within the chosen sub-range to determine the unique deviation coefficients. Once these coefficients are calculated, the sensor can interpolate temperatures across the entire span with high precision.
Choosing a narrower span minimizes the propagation of calibration uncertainties, ensuring that the results remain reliable over time.
Sensor selection
Instrument characteristics must align with the demands of the selected temperature span. Standard platinum resistance thermometers are the primary instruments used to interpolate within these regions. High-temperature thermometers are used for upper spans, while capsule-type sensors are preferred for cryogenic intervals.
The chosen sub-range determines the mechanical design and the purity of the platinum wire required.
Accuracy limit
The precision of the interpolation is limited by the non-repeatability of the thermometer and the uncertainties of the fixed points. Deviation from the true thermodynamic temperature increases as the operating point moves further from the calibration points. This sub-range interpolation remains the most accurate method available for disseminating the temperature scale.