Acquisition Architecture
Signal measurement techniques for high source impedances prevent the sensor from being loaded by the instrument. In high-impedance sensing, the input resistance of the voltmeter or amplifier is several orders of magnitude higher than the source. This ensures that the voltage divider effect between the source and the meter does not introduce large gain errors.
Input Loading
Every measurement pulls a small amount of current from the source which causes a drop across the internal resistance. With high-impedance sensing, this current is minimized to the femtoampere range to preserve the original signal level. A loading error of one percent occurs when the meter impedance is only one hundred times the source impedance.
Therefore, a source with ten gigohms of resistance requires a meter with at least one teraohm of input resistance. The use of high-impedance sensing is fundamental in the characterization of semiconductor devices and dielectric materials where the signal levels are very small.
Noise Susceptibility
High resistance nodes are prone to picking up capacitive noise and electrostatic interference from the environment. Proper high-impedance sensing requires the use of shielded cables and enclosures to block external fields. Thermal noise also increases with the square root of the resistance, setting a fundamental limit on the resolution of the measurement.
Filtering the signal at the output stage helps to remove high frequency noise without affecting the dc accuracy.
Reference Condition
Calibration requires a laboratory environment with controlled humidity. Drift in the input bias current alters the zero point. Verification against a known high resistance standard confirms the linearity of the scale.