Resistance Measurement
High-accuracy determination of low electrical resistance utilizes separate pathways for current supply and voltage measurement to eliminate lead resistance errors. This technique is designated as four-terminal sensing, which enables precise milliohm measurements on sensors and strain gauges.
Current Separation
Injecting a known current through the outer pair of leads creates a voltage drop across the resistance under test. A high-impedance voltmeter connected across the inner pair measures this voltage without drawing current. Because almost no current flows through the measurement leads, the resistance of these leads does not affect the calculation.
This makes four-terminal sensing ideal for low-value resistance elements like copper shunts or platinum RTDs.
Lead Elimination
Calibration routines use reference resistors with known values to verify the elimination of parasitic lead effects. Thermal electromotive forces can still introduce errors if temperature gradients exist across the connections. Reversing the current direction and averaging the readings removes these thermal voltages.
This ensures that four-terminal sensing achieves high precision even in thermally unstable environments.
Accuracy Limit
Mechanical alignment of the probes determines the physical boundary of the measured resistance region. Slight shifts in probe positioning change the effective path length of the current, which alters the measured value. Contact resistance variations are ignored, but the physical stability of the contact points is still required for measurement consistency.