Vacuum Adjustment
Correction applied to pressure readings when a temperature gradient exists between a vacuum gauge and the volume being measured accounts for the difference in particle density. The thermomolecular pressure correction becomes necessary when the mean free path of the gas molecules is large compared to the diameter of the connecting tube. This effect is most pronounced in the transition and molecular flow regimes.
Physical Principle
Molecules moving from a hot region to a cold region exert a different force than those moving in the opposite direction. Without a thermomolecular pressure correction, the gauge indicates a pressure that is higher or lower than the actual pressure in the chamber. The ratio of the pressures depends on the square root of the ratio of the absolute temperatures.
Knudsen numbers determine the exact formula used for the calculation.
Error Mitigation
Using large diameter tubes reduces the magnitude of the thermal transpiration. Application of the thermomolecular pressure correction is standard practice in cryogenic vacuum systems and high altitude flight instruments. Calibration of the sensor must be performed at the same temperature as the process to avoid the need for this adjustment.
Precision Limit
Uncertainties in the temperature profile along the connecting line affect the final result. The thermomolecular pressure correction assumes a steady state condition that may not exist during rapid pumping. Reliable data requires a stable thermal environment across the entire measurement path.