Humidity Metrology
Sensor validation establishes the relationship between the moisture readings of an instrument and a known reference environment. The relative humidity calibration achieves this by exposing the sensor to controlled humidity levels generated by saturated salt solutions or a two-pressure generator. This ensures that the instrument provides reliable measurements across its operating range, preventing errors in applications such as food storage or semiconductor manufacturing.
Calibration Environment
Standard laboratory procedures require stable temperature conditions because relative humidity is highly temperature-dependent. The relative humidity calibration must be conducted in a draft-free environment with a temperature stability of a fraction of a degree. Even a small temperature variation can cause a significant change in the local relative humidity.
Standard Reference
Salt solutions generate precise relative humidity levels at specific temperatures, providing a simple reference. The relative humidity calibration utilizes these chemical properties to verify the sensor at multiple points. After the sensor is exposed to the environment, the difference between the sensor output and the reference value is recorded.
Calibration Drift
Contamination of the sensor surface can cause the measurement to drift over time. The relative humidity calibration should be performed at regular intervals to detect and correct for this drift. If the drift exceeds the specified tolerance, the sensor must be cleaned or replaced.