Angular Precision
Centrifugal and angular velocity sensors require a precise orientation of their sensitive axes relative to the rotational center of a test platform. Rate table alignment ensures that the input axis of a gyroscope or inertial sensor is perfectly parallel or perpendicular to the spin vector. Misalignment introduces cross-axis sensitivity where the sensor responds to rotations it should ignore.
Precision machining of the mounting surface provides the first level of control.
Orthogonality Verification
Optical autocollimators and mirrors verify the angular position of the test fixture. During rate table alignment, the operator adjusts the tilt of the sensor mount until the reflected beam remains stationary through a full rotation. This process identifies the wobble and runout of the table bearing.
Dynamic Error
Centripetal acceleration at high rotation speeds can induce structural deflections in the mounting arm. Such deformations during rate table alignment change the effective orientation of the sensor under test. High speed cameras or laser interferometers monitor these shifts to provide real time corrections.
The software then applies a compensation matrix to the sensor output based on the measured deflection at different angular rates.
Fixture Calibration
Thermal expansion of the table shaft shifts the vertical position of the sensor. Stable room temperatures are necessary for sub-arcsecond accuracy.