Compensation Coefficient
Multiplicative constants applied to raw sensor data adjust for the varying sensitivities of a vacuum gauge to different gas species. Utilizing partial pressure correction factors allows a total pressure gauge to provide accurate readings when the gas composition differs from the calibration gas, which is usually nitrogen. These values are derived from the relative ionisation cross sections or thermal conductivities of the gases involved.
Gauge Application
Pirani gauges and hot cathode ionisation gauges require specific adjustments because their physics are inherently gas dependent. For example, applying partial pressure correction factors when measuring helium prevents an underestimation of the pressure by a factor of six. Capacitance manometers are the only common vacuum instruments that do not require these multipliers because they measure force directly.
Metrological Derivation
Laboratory testing against a primary standard establishes the ratio between the true pressure and the indicated pressure for various gases. These partial pressure correction factors are published in manufacturer data sheets and must be programmed into the vacuum controller. Accuracy is maintained by ensuring the gas mixture remains stable and the gauge is operating within its linear range.
Process Safety
Incorrect values can lead to dangerous over pressurisation or failure of the vacuum process.