Analytical Parameter
Theoretical frequency calculations establish the boundary where the phase and amplitude of eddy currents in a cylindrical conductor become non-linear. The limit frequency is a calculated value based on the diameter, electrical conductivity, and magnetic permeability of the specimen. It serves as a reference point for selecting the optimal operating frequency for tube and bar testing.
Frequency Selection
Operating frequencies are chosen as a ratio of this calculated value to optimize sensitivity for different defect types. To detect outer-surface defects, the instrument frequency is typically set several times higher than the limit frequency. Conversely, lower ratios are used when inspecting for internal or sub-surface flaws to maximize depth penetration.
This selection balances the trade-offs of the skin effect.
Mathematical Model
The calculation uses a standard formula derived from Maxwell equations for electromagnetic induction in a cylinder. This formula requires precise values of material properties which can vary slightly along the length of a manufactured product. Because of these variations, the calculated value is treated as an approximation rather than an absolute threshold.
Reference Block
Calibration standards for bar testing are machined with flat-bottom holes to verify the selected frequency. These holes simulate subsurface defects at specified depths. This setup allows the operator to verify that the phase separation is sufficient for detection.