Parasitic Property
Electrical conductors connecting a sensor to a measurement instrument introduce unwanted impedance into the circuit.
Compensation Technique
Bridge circuits or four wire Kelvin connections remove the effect of cable impedance from the final reading. In a three wire configuration, lead wire resistance is assumed to be equal across all conductors so that the bridge can balance out the additional voltage drop. This requires all wires to be of identical material and length and kept at the same temperature.
Compensation effectively nullifies the parasitic voltage drop that would otherwise be interpreted as a physical change at the sensor.
Material Influence
Resistivity and the temperature coefficient of the conductor determine the magnitude of the error. When lead wire resistance is high, small changes in the ambient environment around the cable can be mistaken for changes in the sensor signal. Using silver or gold plating reduces contact resistance at terminals but does not eliminate the bulk resistance of the wire.
Measurement Error
Total uncertainty in a sensing system includes the uncompensated portion of the cable impedance. If lead wire resistance is not accounted for in a typical two wire 100 ohm platinum sensor, a single ohm of wire resistance results in a temperature error of nearly 2.6 degrees Celsius. Long cable runs in industrial plants exacerbate this effect unless digital transmitters are used close to the sensing element.