Force Analysis
Measurement process capturing rapidly changing loads where the frequency content of the signal is informative. This dynamic force measurement requires sensors with high natural frequencies to avoid signal distortion during impact or vibration testing. Standard static load cells often fail to provide accurate data when the rate of force application exceeds the bandwidth of the instrument.
Signal Acquisition
Piezoelectric sensors are the primary instruments for these high speed applications due to their high stiffness and rapid response times. They generate a charge proportional to the applied mechanical stress, which a charge amplifier then converts into a measurable voltage. Signal conditioning must account for the time constant of the electronics to prevent drift during longer duration events.
Calibration Standard
Secondary calibration methods involve comparing the output of the device under test against a reference transducer using a shaker table or a drop weight system. Traceability is maintained through comparison with standards defined by national metrology institutes. Factors such as the mass of the mounting hardware and the stiffness of the contact surface introduce measurement errors that must be compensated for in the final calculation.
Accuracy Limit
Harmonic distortion occurs when the excitation frequency approaches the resonance of the sensor assembly. Users specify a safe operating range, typically up to twenty percent of the natural frequency, to ensure linear behavior.