Instrument Standardization
The systematic alignment of an instrument or sensor output wherein the reference values are adjusted specifically at the zero and full-scale endpoints establishes a standardized linear response across the entire operating range. During terminal-point calibration, the output curve is forced to pass exactly through these two terminal points, eliminating end-point errors. This method provides a direct, practical baseline for calculating the linearity of the measurement device.
By locking the two extremes of the range, the technician ensures that any intermediate drift can be easily measured and quantified.
Mathematical Correction
Slope and offset adjustments are calculated from the measurement values obtained at the extreme limits of the sensor range. In implementing terminal-point calibration, the resulting calibration line becomes the reference against which all intermediate measurements are evaluated. This approach simplifies the calibration process by avoiding complex polynomial curve fitting.
Linearity Evaluation
Deviation of the actual sensor output from the straight line connecting the end points represents the linearity error of the device. When terminal-point calibration is used, this maximum deviation usually occurs in the middle of the sensor range. This distribution of error must be documented to ensure the device operates within its specified accuracy limits.
Recalibration Frequency
Environmental drift or component aging necessitates periodic checking of the end-point responses to maintain accuracy. If terminal-point calibration reveals a noticeable shift in either the zero-point or span value, the sensor must be adjusted to bring it back into compliance. Regular scheduling of this maintenance procedure ensures the reliability of industrial process controls.