Sensing Limit
Inductive sensor performance depends on the relationship between the physical thickness of a metal target and the penetration depth of the electromagnetic field. The target thickness ratio must be kept high enough to prevent the field from penetrating through the target into the backing material. When this ratio falls below a critical threshold, the output becomes sensitive to changes in the substrate composition rather than target position alone.
Substrate Shielding
Thick targets block the field completely and ensure that the sensor response is determined solely by the target distance. In contrast, thin targets allow the field to interact with materials located behind the target, generating severe measurement error. For aluminum and copper targets, a ratio of at least three is standard to eliminate this interference.
Frequency Adaptation
Higher operating frequencies reduce the skin depth and therefore increase this ratio for thin foil targets. This adjustment allows the sensor to operate accurately on thin materials without requiring physical changes to the setup. The selection of the operating frequency is governed by the minimum thickness of the target.
Calibration Reference
Certified standards of the exact target thickness must be used to calibrate the displacement system. This step ensures that the instrument behaves linearly across the full measuring range.