Thermal Transfer
Electrical current flowing through a resistive temperature sensor generates internal heat that raises the temperature of the sensing element above the ambient level. This phenomenon of self-heating dissipation alters the accuracy of temperature measurements by introducing a positive thermal offset.
Measurement Error
The magnitude of the thermal rise depends directly on the excitation current and the thermal resistance of the environment. In stagnant air, the dissipation factor is low, which leads to a higher temperature rise than in moving liquid. This difference means that the same sensor will report different values depending on the flow rate of the surrounding medium.
Mitigation Method
Reducing the excitation current minimizes the internal heat generation but also decreases the signal output. Engineers often choose a pulsed current technique to limit the active time of the measurement and allow the sensor to cool between readings. This approach maintains a high signal to noise ratio without causing a drift in the measured temperature.
Sensor Evaluation
Quantifying the dissipation constant involves measuring the sensor output at two different excitation currents while maintaining a stable bath temperature. The difference in measured resistance reveals the thermal rise per milliwatt of power. This parameter is used to calibrate the measurement software, which applies a mathematical correction based on the applied power.
This calibration is performed during the final production run of the instrument to ensure that the correction is accurate for the specific housing style.