Primary Metrology
Determination of hydraulic or pneumatic pressure through the application of known masses to a piston of precisely measured area establishes a fundamental reference. Regular pressure balance calibration ensures that the effective area of the piston cylinder assembly remains within the specified tolerance. This process links industrial sensors to national standards.
Correction Factor
Local gravity and air buoyancy alter the downward force exerted by the weights. During a pressure balance calibration, the operator applies corrections for the thermal expansion of the steel and the elastic distortion of the cylinder under load. Fluid head corrections must also be calculated to account for the height difference between the reference level and the gauge under test.
Surface tension of the lubricating oil adds a small but measurable force at the piston interface.
Uncertainty Source
Piston tilt and rotation speed influence the stability of the pressure generated. Results from a pressure balance calibration are sensitive to the verticality of the mounting post. Friction between the piston and cylinder introduces non repeatability if the assembly is not cleaned to laboratory standards.
Traceability Chain
Cross floating compares two systems to verify the consistency of the effective area. Documentation of a pressure balance calibration includes the certificate of the mass set and the dimensional measurement of the piston diameter. These records provide the evidence of accuracy required for ISO 17025 accreditation.