Verification Method
A dynamic testing procedure evaluates the response time of a temperature sensor by suddenly moving it from one thermal environment to another. During a plunge testing sequence, the probe is transferred from ambient air into a controlled, circulating bath of water or oil. This action creates a step-function change in temperature that reveals the thermal lag of the instrument.
The results are used to calculate the time constants that govern the behavior of the sensor in real-world applications.
Mechanical Setup
The apparatus used for this test must ensure a consistent velocity and immersion depth. Effective plunge testing requires that the sensor enters the medium at a right angle to the flow.
Data Capture
Electronic logging systems record the voltage or resistance of the sensor at high frequencies. An analyst looks at the plunge testing data to find the point where the signal reaches specific percentages of the total change. These data points are then used to model the transfer function.
Precise timing is required to isolate the moment of impact.
Field Correlation
The values obtained in the lab provide a baseline for the performance of the probe. Because real-world conditions vary, plunge testing serves as a repeatable way to compare different brands or designs. It confirms that the internal construction of the sensor is sound.