Weighting Factor
Coefficient partitioning the temperature dependence and the structural dependence of molecular relaxation times. In thermal analysis, the non-linearity parameter x determines how much the current structural state influences the rate of relaxation compared to the actual temperature. It ranges from zero to one.
Energy Partition
A value of one indicates that the material relaxation depends only on the temperature. A value of zero suggests that the internal structure alone controls the process. Most glassy materials fall between these extremes, showing a complex mix of both factors.
Kinetic Balance
The parameter appears in the Tool-Narayanaswamy-Moynihan equation to describe why the relaxation slows down as the glass becomes denser. This effect creates the asymmetrical peak seen in heat capacity measurements during heating. Determining the correct value is necessary for predicting the long-term stability of amorphous pharmaceuticals and polymers.
Fit Accuracy
Calibration of this parameter requires a series of aging experiments at temperatures just below the glass transition. The analyst compares the recovered enthalpy after different storage times to the model predictions. Errors in the value lead to poor predictions of the physical aging rate.
Verification involves using a double-scan technique where the second scan provides a baseline for the first. The resulting value must be consistent across different cooling and heating cycles to be considered valid for the material.