Sensing Architecture
Pressure sensing instruments rely on physical displacement between parallel conductive surfaces to alter electrical storage capacity. A ceramic capacitive transducer uses an aluminum oxide diaphragm positioned across a narrow gap from a rigid substrate containing screen-printed electrodes. Deflection of the flexible element changes capacitance in direct relation to applied force.
The dry measurement cell isolates the medium from internal electronics, preventing fluid contamination across industrial process lines. Mechanical stops behind the diaphragm protect the assembly against hydraulic shock up to four times the nominal rating. This structural limit defines the overpressure boundary where elastic recovery yields to permanent sensor damage.
Dielectric Behaviour
High purity alumina provides chemical resistance and high mechanical stiffness, preventing hysteresis during pressure cycling. Deflection remains below two micrometres across full scale pressure ranges, maintaining a predictable gap distance. Parasitic capacitance within wiring networks introduces signal bias that requires active guarding techniques.
Thermal Response
Ambient temperature fluctuations alter diaphragm dimensions and modify the dielectric constant of the internal cavity gas. Thermal expansion coefficients between the ceramic body and glass seal match closely to limit zero drift across operating bands. Uncompensated shifts appear as offset changes rather than span errors.
Calibration Boundary
Factory verification tests output linear response against deadweight testers under controlled atmospheric conditions. Field installation effects like mounting torque can distort the sensor frame, altering baseline capacitance prior to zero adjustment. Recalibration requires resetting baseline values without altering the factory span coefficient.