Acquisition Interface
Electronic conditioning blocks design-engineered for physiological voltage measurements form the initial stage of clinical recording systems. The biopotential instrumentation front end must capture sub-millivolt signals from the skin surface while rejecting large common-mode voltages and high-frequency interference. To prevent signal attenuation, each input channel possesses an exceptionally high input impedance of at least one gigaohm.
Common Rejection
Differential amplification isolates the minute physiological voltages from surrounding electrostatic noise. A high common-mode rejection ratio of over one hundred decibels is typical in a standard biopotential instrumentation front end. Active circuitry like a right-leg drive circuit dynamically drives the subject with a compensating signal to reduce the common-mode voltage further.
Noise Performance
Thermal and electronic noise within the semiconductor components determines the ultimate resolution of the system. The intrinsic noise of a biopotential instrumentation front end is constrained by low-noise operational amplifiers. In the band from point one hertz to one hundred hertz, the noise must remain below one microvolt peak-to-peak.
Maintaining this level ensures that microvolt-level features like the cardiac QRS complex are resolvable.
Isolation Barrier
Patient safety requires galvanic separation from the power grid. Optical or capacitive isolation barriers separate the biopotential instrumentation front end from the main digital processor. A medical-grade isolation barrier limits the leakage current flowing through the patient to less than ten microamperes during any fault condition.