Orientation Sweep
Evaluating multi-axis accelerometers requires rotating the device through a sequence of static positions relative to the gravity vector. Performing a tumble test allows the calibration system to capture the sensor response at multiple angles to determine scale factors and offsets. This procedure ensures that the accelerometer provides accurate readings regardless of its physical orientation.
Gravity Calibration
The local gravitational acceleration provides an incredibly stable and highly accurate reference standard for the testing cycle. By measuring the sensor output when aligned both parallel and anti-parallel to gravity, the algorithm isolates the zero-g offset from the sensitivity scale. This method requires no external electrical or magnetic references.
Error Isolation
Dynamic effects during the transition between test positions can introduce transient noise and skew the static measurements. Software routines must wait for the sensor signal to stabilize after each rotation before capturing the reference data. This delay prevents mechanical vibrations from corrupting the calibration calculation.
Mechanical Execution
High-precision rotary stages execute the angular movements with micro-radian repeatability to prevent positioning errors from influencing the results. Even minor misalignments of the test fixture relative to the true vertical axis can introduce projection errors into the collected data. In industrial production lines, automated pick-and-place arms position the devices on the rotary stage to ensure high throughput while maintaining the precise alignment needed for accurate measurement.
This automation keeps the test cycle both fast and repeatable.