Sensor Instability
A permanent change in the output of a measurement device occurs when internal materials expand or contract at different rates during temperature changes. This thermal stress drift results from the mechanical tension exerted by the package on the sensing element. Unlike reversible temperature coefficients, this phenomenon often appears as a non-return to the original zero point.
Temperature Cycle
Mismatched coefficients of thermal expansion between the silicon die and the ceramic substrate create a bending moment. As the assembly cools, thermal stress drift accumulates if the die attach material does not provide sufficient compliance.
Package Strain
Engineers evaluate the magnitude of the error by subjecting the component to repeated cycles between its storage limits. Measuring the thermal stress drift involves logging the output at a reference temperature after each thermal excursion to track the shift. High-precision resistors and voltage references are particularly susceptible to this type of aging.
Mitigation strategies include the use of soft adhesives or lead-frame designs that isolate the die from the external package. Characterizing this behaviour is essential for instruments that must maintain their accuracy over years of field service in harsh environments.
Recovery Limit
Calibration intervals are often determined by the rate at which the thermal stress drift moves the device out of its specified tolerance. If the drift exceeds the allowed limit, the module must be re-aligned or replaced to ensure the integrity of the data.