Hysteresis Analysis
Testing procedures that quantify the difference in sensor output at a specific temperature when approached from above and below reveal the stability of the mechanical assembly. Performing a thermal hysteresis measurement involves heating the device to its maximum limit and then cooling it back to its minimum limit. The discrepancy in the readings at the midpoint of these cycles defines the hysteresis value.
This metric is usually recorded after the sensor has undergone several full temperature cycles to ensure the internal stresses have reached a steady state.
Physical Origin
Internal stresses in the housing or the adhesive used to mount the sensor chip cause thermal hysteresis measurement results to deviate from zero. When materials with different expansion coefficients are bonded together, they may not return to their original state immediately after a thermal cycle. This lag creates a non-repeatable error that complicates the calibration process.
Quantification Method
A typical thermal hysteresis measurement requires multiple temperature cycles to ensure the data is statistically valid. The sensor is held at each soak point until the internal temperature stabilizes to avoid transient errors. The maximum difference between the up-ramp and down-ramp readings is reported as a percentage of the total span.
Impact on Accuracy
High values found during thermal hysteresis measurement indicate that the sensor may not provide consistent data in environments with fluctuating temperatures. This error is particularly difficult to compensate for because it depends on the thermal history of the device and is not determined by the current temperature alone. Minimizing this effect requires careful selection of materials and assembly techniques.