Reference Standard
Precision temperature measurement relies on high-purity platinum wire to realize the fixed points defined by the International Temperature Scale. SPRT thermometry uses a strain-free winding of platinum that changes its electrical resistance in a highly predictable manner when exposed to heat. These instruments are the primary transfer standards between national metrology institutes and calibration laboratories.
They provide the most accurate means of measuring temperature between the triple point of hydrogen and the freezing point of silver.
Resistance Ratio
Calibration of the probe involves measuring its resistance at specific physical constants, such as the triple point of water. In SPRT thermometry, the fundamental parameter is the ratio of the resistance at a given temperature to the resistance at the water triple point. This ratio, known as W, removes the dependence on the absolute resistance value of the wire and accounts for minor variations in the manufacturing process.
A high W value at the gallium melting point indicates that the platinum is of sufficient purity for primary standard use.
Self-Heating
Current passed through the wire to measure its resistance generates a small amount of internal heat that raises the temperature of the sensor. SPRT thermometry requires measurements at two different current levels to calculate and remove this error. By extrapolating the results to zero current, the actual temperature of the environment is determined.
This correction is necessary when measuring in low-thermal-conductivity environments like stagnant air or vacuum.
Calibration Traceability
Maintaining the accuracy of the probe requires periodic verification against known physical fixed points. Over time, mechanical shocks or chemical contamination can introduce defects into the crystal lattice of the platinum, causing the resistance to drift. SPRT thermometry remains the gold standard because these changes can be detected by small shifts in the triple-point resistance.
Properly handled, a standard thermometer maintains its calibration for several years of laboratory service.