System Characterization
Deformation of the testing machine frame and fixtures under load creates a discrepancy between the measured and actual specimen displacement. Through instrument compliance mapping, engineers determine the spring constant of the instrument assembly across the operating temperature range. This step is necessary to prevent the stiffness of the machine from corrupting the specimen data.
Correction Execution
Raw displacement data are adjusted by subtracting the machine deflection calculated from the measured force and the compliance map. During instrument compliance mapping, a highly rigid steel or ceramic blank is compressed or stretched to record the absolute deformation of the system. This baseline deflection is then used by the software to apply a real-time correction to each measurement point.
Accurate subtraction yields the true strain of the polymer specimen.
Calibration Standard
Rigid metallic samples with known elastic properties provide the standard for verifying the compliance map. System verification runs must show that the corrected stiffness of the test bar matches its theoretical value. Any remaining deviation is used to adjust the compliance coefficient.
This calibration occurs during annual maintenance or when changing fixture configurations.
Mechanical Boundary
Extremely stiff polymer samples measured at cryogenic temperatures approach the stiffness limit of the test fixtures. In this regime, instrument compliance mapping becomes necessary because the fixture deflection can exceed the sample deformation. This limit defines the maximum specimen stiffness that the instrument can accurately resolve.