Shielding Strategy
Active shielding techniques that drive a conductive enclosure or trace to the same potential as the signal conductor eliminate parasitic current paths in high sensitivity measurements. By maintaining a zero potential difference between the signal and its surroundings, dynamic guarding prevents charge from flowing through insulation resistance. This method is used in electrometers where picoampere currents are measured.
Stray capacitance of the system is transformed into a negligible load.
Leakage Suppression
The guard amplifier follows the input signal with precision to ensure the shield tracks the conductor. Any voltage offset in this amplifier appears as a leakage source. High gain and low offset are required for the guard driver.
Frequency Limitation
Bandwidth restrictions in the guard driver limit the effectiveness of the shield at high frequencies. When the guard signal lags the primary signal, the guarding effect diminishes and the input capacitance appears to increase. This phase lag can introduce noise.
System Verification
Testing of the guard efficiency involves measuring the input current with the guard enabled versus disabled. A successful implementation shows a reduction in settling time and noise. Calibration protocols require checking the guard gain and offset periodically.
Mechanical integrity of the shields remains necessary for the measurement.