Spectral Mapping
Frequency adjustment of a spectrometer scale aligns the detector pixels with known spectral reference lines to establish a coordinate system. Executing a dispersion calibration ensures that the observed signals appear at their correct positions on the horizontal axis. This step is required for any instrument where the relationship between position and energy is non linear.
Reference Emission
Standardized light sources like argon or mercury lamps provide a series of sharp lines at precisely known wavelengths. During dispersion calibration, the software records the pixel location of each line to build a mathematical model of the optical system. A polynomial fit typically describes the mapping between the sensor array and the spectral units.
Accuracy Check
Thermal expansion of the grating or the housing can cause the spectral scale to shift or stretch over time. Periodic dispersion calibration compensates for these mechanical changes and maintains the validity of the data. High precision systems use a secondary standard like a laser or a neon lamp to verify the scale during every measurement.
Reliability of the reported wavenumbers depends on the stability of this mapping.
Calibration Drift
Mechanical shocks or significant temperature swings are the primary causes of a loss in spectral alignment. A failed dispersion calibration results in incorrect identification of chemical species or erroneous stress measurements in materials. Stability of the mount and the detector determines the required frequency for this maintenance task.