Electrostatic Protection
Low-current measurement configurations require active electrostatic isolation to prevent leakage currents from distorting sub-picoampere signals. Implementing triaxial guard shielding involves a cable structure with an inner force conductor, an inner guard shield driven at the same potential, and an outer chassis ground shield. This arrangement eliminates the potential difference between the force conductor and its immediate shield, reducing leakage to near zero.
Guarding Mechanism
Active guarding works by utilizing a voltage follower circuit to drive the guard shield at the exact potential of the signal conductor. Because there is no voltage drop across the dielectric material separating them, no current can flow through the insulation resistance, regardless of how high that resistance is. This technique effectively bypasses the cable capacitance, which dramatically decreases the settling time of the measurement circuit.
It is particularly effective when measuring high-resistance materials or the tiny currents produced by photodetectors. The outer shield remains at ground potential to protect the guard loop from external electromagnetic interference.
Installation Constraint
Cable selection and connection routing must maintain the isolation of the guard conductor from the ground return throughout the signal path. Specialized triaxial connectors are required to prevent accidental shorting of the high-voltage guard.
Calibration Verification
Annual instrumentation checks verify the isolation resistance between the three conductor paths using a specialized electrometer. This verification ensures that humidity or dust has not degraded the cable shielding performance.