Signal Buffer
Active circuits drive the inner shield of a specialized cable to the same potential as the signal wire. Use of triaxial cable drivers eliminates the leakage current between the signal conductor and the first shield. This is necessary for measuring currents in the picoampere range where the insulation resistance of standard coaxial cable is insufficient.
Guard Potential
The driver provides a low impedance source that tracks the input voltage with high precision. In a triaxial cable, the signal travels on the center conductor and the guard potential is applied to the inner shield. Because there is no voltage difference between the center and the shield, the leakage current is zero.
The outer shield remains at ground potential to provide protection from external electromagnetic fields.
Settling Time
Charging the capacitance of a long cable requires a driver with sufficient current capability and bandwidth. Triaxial cable drivers must settle to the final value quickly to allow for high speed measurements. The effective capacitance of the cable is reduced because the driver maintains the shield at the same potential as the signal.
This reduction in the rc time constant allows for faster transitions than would be possible with an unguided cable.
System Integration
Connection of the driver to the cable requires specialized connectors that maintain the isolation between the conductors. Any contamination at the connector interface introduces leakage paths that bypass the guard. Verification of the system performance involves measuring the noise floor with the cable attached.
Periodic calibration of the driver offset ensures that the guard potential remains exactly equal to the signal voltage.