Voltage Generation
Unintentional voltage is produced in a measurement circuit when dissimilar metals are joined in the presence of a temperature gradient. This voltage is called thermal offset and it behaves as an unwanted signal that interferes with low-level sensor readings. It occurs most often in thermocouple circuits and high-precision resistance measurements where small voltages are converted into temperature values.
This effect is a primary source of error in precision thermometry.
Measurement Error
Small voltages generated by temperature differences along a signal wire alter the apparent resistance or voltage of the sensor. A thermal offset of just a few microvolts can shift a temperature reading by several tenths of a degree. This error is difficult to distinguish from actual sensor changes during operation.
It requires specialized measurement techniques to detect and isolate.
Calibration Mitigation
High-precision calibration systems use current reversal techniques to cancel out the effects of unwanted voltages. This method eliminates the thermal offset from the measurement calculation by averaging readings taken with opposite current directions.
Prevention Method
Keeping the junctions of dissimilar metals at a uniform temperature minimizes the generation of parasitic voltages. This layout reduces thermal offset and ensures the stability of precision instrumentation. Using continuous cabling without intermediate connections also limits this error.