Calibration Reference
Metrological processes designed to verify the dynamic force or acceleration output of high-frequency vibration tables ensure the traceability of motion measurements. The process of piezoelectric shaker calibration compares the output of a test shaker against a reference accelerometer certified by a national standards laboratory. This comparison establishes the relationship between the applied electrical drive signal and the resulting physical acceleration.
Accurate calibration is necessary for testing seismic sensors and high-performance accelerometers.
Frequency Range
Dynamic verification must cover the full operating band of the instrument. During piezoelectric shaker calibration, the system is swept across a wide range of frequencies, often from a few hertz to several kilohertz, to measure the frequency response of the shaker. This sweep identifies any structural resonances that could affect the accuracy of subsequent sensor testing.
Transducer Sensitivity
The sensitivity of the shaker defines the relationship between input voltage and output acceleration. In piezoelectric shaker calibration, this parameter is measured at a specific reference frequency to provide a baseline for future measurements.
Uncertainty Assessment
Every measurement contains a degree of uncertainty that must be quantified and reported. Determining the uncertainty in piezoelectric shaker calibration involves analyzing contributions from reference sensor calibration, environmental temperature fluctuations, electrical noise, and alignment errors. The final certificate lists the expanded uncertainty with a coverage factor of two, providing a ninety-five percent confidence level for the calibrated sensitivity value.