Extensometer Calibration
Standardized calibration procedures for static uniaxial extensometer systems ensure traceability in mechanical material testing applications. Compliance with ISO 9513 dictates how displacement transducers are classified based on maximum permissible error during gauge length verification. Calibration apparatuses utilize laser interferometers or micrometer heads to apply incremental displacements.
Accuracy classes range from Class 0.2 to Class 2.
Error Boundary
Gauge length determination requires precise physical measurements between knife edges or contact points. Relative bias error and absolute deviation limits govern the assigned accuracy class. Temperature stability during testing must remain within strict tolerances to avoid thermal strain errors.
Verification Protocol
Verification cycles require ten incremental displacement steps applied in both ascending and descending directions. Adherence to ISO 9513 requires reporting the hysteresis, repeatability, and relative error of the extensometer across its full working range. Differential thermal expansion between the calibration rig and the instrument under test causes measurement bias if ambient thermal equilibrium is not maintained.
Hysteresis loops identify mechanical backlash in flexure blades or mounting arms.
Certificate Compliance
Calibration certificates state the verified measuring range and corresponding accuracy class for specific gauge length settings. Recalibration intervals depend on mechanical wear of contact points and operating environment severity. Accredited laboratories issue documentation traceable to national metrological institutes.