Thermal Leakage
Thermal leakage along the sheath of a probe causes a measurement error by pulling the sensing junction away from the process temperature. Stem conduction loss occurs when a temperature gradient exists between the immersion tip and the ambient environment outside the vessel. This effect results in a reading that is lower than the actual process temperature when the sensor is hotter than the surroundings.
Immersion Depth
Increasing the length of the sensor buried within the fluid reduces the magnitude of this error. A common rule of thumb suggests an immersion depth of at least ten times the diameter of the probe. This distance ensures that the sensing element is thermally isolated from the mounting hardware.
Design Mitigation
Using thin walled sheaths or materials with lower thermal conductivity can limit the flow of heat along the stem. Lagging or insulation around the mounting point also helps to maintain the probe body at the process temperature. Engineers must balance the mechanical strength of the thermowell against the need to minimize stem conduction loss in high velocity flows.
Calculating the expected error involves analyzing the thermal resistance of the entire sensor assembly to predict behavior under different gradients.
Measurement Error
Inadequate immersion leads to offsets in industrial thermometers. This error is verified by varying the depth during a field check.