Molecular Dissipation
Time dependent loss of internal stress within a bonded polymer layer occurs when polymer chains reorient to reach a lower energy state. In strain gauge pressure sensors, adhesive relaxation results in a drift in the zero point or a shift in the calibration curve over time. This process typically accelerates with elevated temperatures because thermal energy increases the mobility of the polymer backbone.
Sensor Performance
Creep effects originate from the inability of the bonding agent to maintain a constant strain transfer from the diaphragm to the sensing element. When adhesive relaxation occurs in a pressure transducer, the indicated pressure drops even if the applied physical load remains constant. Engineers quantify this by monitoring the output signal under a static load over several hundred hours to ensure it meets stability requirements.
If the rate of decay is too high, the sensor will fail to maintain its accuracy over its service life.
Thermal Dependency
Stress dissipation rates follow the Arrhenius equation where the rate of change increases exponentially with the reciprocal of absolute temperature. Proper curing cycles reduce adhesive relaxation by ensuring a high degree of cross linking during the initial manufacturing stage. Uncured monomers act as plasticizers that facilitate chain movement and degrade long term stability.
Metrological Bound
Long term stability specifications define the maximum permissible error resulting from polymer creep and other aging factors. Calibration intervals are set based on the predictable decay of internal stresses within the sensor assembly. Once the polymer reaches an equilibrium state, the rate of change diminishes.